Fisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine Test Dipstick CLIA Waived, Model/ Part:FHC-101/5000; Henry Schein One Step+ hCG Urine Strip Test, Model/ Part: FHC-101/9004072; Alere hCG Dipstick (25mIU/mL) in Urine, Model/ Part: FHC-101/92211; Cardinal Health hCG Dipstick Rapid Test CLIA Waived, Model/ Part: FHC-101/B-1077-21; Medline hCG Pregnancy Test Strip, Model/ Part: FHC-101/MPH11025HCG; Pro Advantage by NDC Urine Pregnancy Test Strip, Model/ Part: FHC-101/P080023;
Class II · Ongoing
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Advertisement
Advertisement
Affected lots / codes — check yours against thisAll lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900527/00613647000588/1155811305, 2018-01-15; 5000/20612479202751/40008059, V05; 9004072/10304040024093/1155800504, 2014-06-09; 92211/10815845020093/40008089, 05/25; B-1077-21/20885380020244/40008054, 04/2025; MPH11025HCG/10884389191873/40006717, 6/2025; P080023/M220P0800233/40008087, 2025-05-09
- Recalling firm
- Alere San Diego, Inc., San Diego, CA
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Ongoing
- Quantity
- 13,256,075
- Distribution
- Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ire…
- Recall initiated
- August 20, 2026
- Classified
- September 25, 2026
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-3273-2026%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-3273-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →