FDA Medical Device · Z-3188-2026

Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0

Class II · Ongoing

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

Advertisement
Advertisement
Affected lots / codes — check yours against thisAll Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)
Recalling firm
Boston Scientific Corporation, Marlborough, MA
Classification
Class II (temporary or reversible harm possible)
Status
Ongoing
Quantity
982 units
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and th…
Recall initiated
July 22, 2026
Classified
September 16, 2026
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-3188-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →