Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
Class II · Ongoing
Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.
Advertisement
Advertisement
Affected lots / codes — check yours against thisModel Number: INFX-8000C/B3 UDI code: Pending SID Number:30101165 Serial Number: B3E2422005 Model Number: INFX-8000F/SG UDI code: Pending SID Number: 30093711 Serial Number: SGC2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30029686 Serial Number: SZA2422001 SID Number: 79000002 Serial Number:SZA2442004 Model Number: INFX-8000V/SX UDI code: Pending SID Number: 30049491 Serial Number: SXA2432001 SID Number: 30100053 Serial Number: SXA2432002 Model Number: INFX-8000V/B9 UDI code: Pending SID Number: 30059175 Serial Number: B9F2412001 Model Number: INFX-8000V/ST UDI code: Pending SID Number: 30059180 Serial Number: STA2412001 Model Number: INFX-8000V/SZ UDI code: Pending SID Number: 30095017 Serial Number: SZA2432003 SID Number: 30097052 Serial Number: SZA2422002
- Recalling firm
- Canon Medical System, USA, INC., Tustin, CA
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Ongoing
- Quantity
- 10 systems
- Distribution
- U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
- Recall initiated
- August 4, 2026
- Classified
- August 25, 2026
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-3014-2026%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-3014-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →