FDA Medical Device · Z-3012-2026

Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

Class II · Ongoing

A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.

Advertisement
Advertisement
Affected lots / codes — check yours against thisLot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901
Recalling firm
Midmark Corporation, Versailles, OH
Classification
Class II (temporary or reversible harm possible)
Status
Ongoing
Quantity
3624 units
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall initiated
July 20, 2026
Classified
August 25, 2026
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-3012-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →