FDA Medical Device · Z-2927-2026

Thermo Fisher Diagnostics Neisseria meningitidis Gp C 2ml/VL, Product Number R30166901; invitro diagnostics

Class II · Ongoing

Certain lots of Neisseria meningitidis Gp C 2ml/VL, R30166901, are not performing to IFU criteria not positive to ATCC13102 within 60 Seconds.

Advertisement
Advertisement
Affected lots / codes — check yours against thisUDI/DI 05056080501161, Lot Numbers: 6214791, exp. 18/03/2025; 6175327, exp. 04/12/2027; 6148809, exp. 30/09/2027; 6002010, exp. 31/07/2027.
Recalling firm
Remel Europe Ltd., Dartford,
Classification
Class II (temporary or reversible harm possible)
Status
Ongoing
Quantity
26 units
Distribution
US Nationwide distribution.
Recall initiated
July 17, 2026
Classified
August 7, 2026
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2927-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →