FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).
Class II · Ongoing
Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.
Advertisement
Advertisement
Affected lots / codes — check yours against thisPart Number: NGDS-ASY-0007 UDI-DI code: (01)00851458005136 Lot Numbers: 311825D 312025D 313526D 322925D
- Recalling firm
- Biofire Defense, Salt Lake City, UT
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Ongoing
- Quantity
- 122 kits (3,660 pouches)
- Distribution
- U.S. Nationwide distribution in the states of AL, CA, DC, FL, GA, HI, KY, LA, MA, MD, MN, NC, NH, NV, NY, OH, RI, TN, TX, UT, VA and WA.
- Recall initiated
- July 16, 2026
- Classified
- August 4, 2026
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2877-2026%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2877-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →