FDA Medical Device · Z-2685-2026

ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 14Fr Introducer, 13cm, Long Taper, Introducer Product Code 0052-3052. heart pump and accessories

Class I · Ongoing

A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

Advertisement
Advertisement
Affected lots / codes — check yours against thisPump Set Code 0052-0038; Introducer Product Code 0052-3052. GTIN: 00813502011708, 00813502012729. All current batches in field.
Recalling firm
Abiomed, Inc., Danvers, MA
Classification
Class I (most serious — risk of serious harm)
Status
Ongoing
Quantity
1,540 units
Distribution
Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan …
Recall initiated
June 11, 2026
Classified
July 10, 2026
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2685-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →