FDA Medical Device · Z-2416-2015

ADVIA Chemistry Acetaminophen reagents; for in vitro diagnostic use in the quantitative determination of acetaminophen in human serum and plasma (lithium heparin) on ADVIA Chemistry systems.

Class II · Terminated

A change in the concentration of N-acetylcysteine (NAC) that may cause interference for the ADVIA¿ Chemistry Systems Acetaminophen assay for certain lots.

Advertisement
Advertisement
Affected lots / codes — check yours against thisCatalog Number 07989138, Siemens Material Number (SMN) 10327381, Lot Numbers: 46251, 46564, 46369, 46370, 46619, 46782, 47001, 47157, 47558; with expiration dates June 30, 2015 through April 30, 2016.
Recalling firm
Siemens Healthcare Diagnostics, Inc., Newark, DE
Classification
Class II (temporary or reversible harm possible)
Status
Terminated
Quantity
1032
Distribution
Worldwide Distribution - US (nationwide) and to the countries of : Canada, Netherlands, Great Britain, Norway, Sweden, Spain, Italy, Germany, France, Finland, Belgium, Portugal, Czech Republic, Israel, Poland, Turkey and Bahrain.
Recall initiated
June 30, 2015
Classified
August 19, 2015
Terminated
September 1, 2016
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2416-2015%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →