The Hospira MedNet Medication Management Suite (MMS). Product Usage: The Hospira MedNet Medication Management Suite (MMS) is intended to facilitate networked communication between MMS compatible computer systems and Hospira infusion pumps. The MMS provides trained healthcare professionals with the capability to send, receive, report, and store information from interfaced external systems, and to configure and edit infusion programming parameters. The MMS is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by a trained healthcare professional according to physicians orders.
Class II · Terminated
Hospira MedNet 5.5, 5.8.1, and 5.8.2 contains software defect where the dosing units of "nanog/kg/min" and "milliUnits/min" are not sent to a Plum A+ device (Version 13.40, 13.41, and 13.4.2) via Auto Programming when those dosing units are not properly established in the customized drug library. A dosing unit mismatch of "nanog/kg/min" or "milliUnits/min" on an unmatched medication will produce a
Advertisement
Advertisement
Affected lots / codes — check yours against this1) Software Version 5.5, List Number: 16037-04-13; 2) Software Version 5.8.1, List Number: 16037-04-31; 3) Software Version 5.8.2, List Number: 16037-04-33
- Recalling firm
- Hospira Inc., Lake Forest, IL
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Terminated
- Quantity
- 13 units
- Distribution
- US Nationwide in the states of FL, LA, PA, WA, WY and the country of Hong Kong
- Recall initiated
- June 6, 2014
- Classified
- July 9, 2014
- Terminated
- December 6, 2016
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2027-2014%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-2027-2014%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →