FDA Medical Device · Z-1947-2026

AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).

Class I · Ongoing

Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.

Advertisement
Advertisement
Affected lots / codes — check yours against thisKit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots.
Recalling firm
North American Rescue LLC., Greer, SC
Classification
Class I (most serious — risk of serious harm)
Status
Ongoing
Quantity
21
Distribution
US: CA, WA
Recall initiated
March 27, 2026
Classified
April 29, 2026
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1947-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →