Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test
Class II · Terminated
The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400 plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following statement in the Limitations-interference section: "In very rare cases, gammopathy, in particular type IgM (Waldenstr¿m s macroglobulinemia), may cause unreliable results."
Advertisement
Advertisement
Affected lots / codes — check yours against thisLot codes: 11021500 7/31/2016 12361201 5/31/2017 11031700 9/30/2018 15247301 8/31/2017 11071400 1/31/2016 18036001 11/30/2017 11071600 8/10/2016 19329601 1/31/2018 11101500 1/21/2016 20511301 5/31/2018 11111600 4/30/2018 24076801 8/31/2018 60482201 2/29/2016 60852601 5/31/2016 61385701 9/30/2016 62032101 1/31/2017
- Recalling firm
- Roche Diagnostics Corporation, Indianapolis, IN
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Terminated
- Quantity
- 35,111
- Distribution
- Nationwide Distribution
- Recall initiated
- January 2, 2018
- Classified
- April 25, 2018
- Terminated
- April 25, 2019
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1519-2018%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1519-2018%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →