FDA Medical Device · Z-1472-2023

Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network,

Class II · Ongoing

The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.

Advertisement
Advertisement
Affected lots / codes — check yours against thisUDI-DI: 04049098048989 (M300), 04049098095778 (M300+); All Systems Running Software versions: VG2.4 and lower
Recalling firm
Draeger Medical Systems, Inc., Andover, MA
Classification
Class II (temporary or reversible harm possible)
Status
Ongoing
Quantity
19,429 systems
Distribution
Worldwide - US Nationwide distribution.
Recall initiated
April 5, 2023
Classified
April 27, 2023
Initiation
Voluntary: Firm initiated