Draeger Infinity M300 and M300+, for use with the ICS to monitor ECG and pulse oximetry on ambulatory and non-ambulatory adult and pediatric patients using wireless communication over the Infinity patient monitoring network,
Class II · Ongoing
The software on the Infinity CentralStation drops peaks on narrow waveforms causing the Infinity M300 to fail to meet Frequency and Impulse Response requirements. In this condition, ECG waveforms with unusually narrow and/or high frequency QRS may be displayed or printed with QRS amplitudes intermittently lower than actual.
Advertisement
Advertisement
Affected lots / codes — check yours against thisUDI-DI: 04049098048989 (M300), 04049098095778 (M300+); All Systems Running Software versions: VG2.4 and lower
- Recalling firm
- Draeger Medical Systems, Inc., Andover, MA
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Ongoing
- Quantity
- 19,429 systems
- Distribution
- Worldwide - US Nationwide distribution.
- Recall initiated
- April 5, 2023
- Classified
- April 27, 2023
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1472-2023%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1472-2023%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →