FDA Medical Device · Z-1455-2025

Brand Names: MyDay Toric, LensCrafters One Day Premium Daily Disposable Contact Lenses For Astigmatism, Pearle Vision 1 Day Premium For Astigmatism, REVEAL 1-DAY TORIC, Vision Source Reveal Toric, Voyant 1-Day Premium Toric. Brand Names (OUS only): CADENCE DELUXE 1-DAY TORIC, DUBLAN PREMIUM OXYGEN 1 DAY TORIC, EV LINARIAL TORIC (30), EYEXPERT PURE ASTIGMATISM (30), LENTI!!AS A ASTIGMATISMO 30PK, LENTSOFT DIARIA CONFORT+ SILICONA ASTIGMATISMO, MULTILENS CONTACT ELITE SILICONE 1 DAY TORICS, OPTIONS SUPREME 1 DAY TRC (30), PRIMA + 1 DAY ASTGM (30), PROTECT PLATINUM FOR ASTG (30), SOL- LENS 1 DAY SILICONE PREMIUM TORIC, UGO LENS DAY ADVANCED TORICA, WHITE BOX STENFILCON A TORIC 07536, MEDIFLEX SUP 1D TORIC, Kirkland Signature Toric Premium Daily Disposable, Specsavers easyvision Linarial Toric. Product Name: (Stenfilcon A) Contact Lens Model/Catalog Number: MyDay Toric Product Description: 54% water content, Toric, cast moulded Silicone Hydrogel, In-monomer tinted, contact lens, for Daily Disposable Wear. Blister packaged. Component: No

Class II · Ongoing

a limited number of lots were manufactured with an incorrect cylinder power.

Advertisement
Advertisement
Affected lots / codes — check yours against thisBlister Lot Numbers: (1) 6461501582, (2) 6461501590, (3) 6448502235, (4) 6472500679. Carton Lot numbers: (1) 19346502595028, (2a) 344446500117, (2b) 19346502613, (2c) 21455502575, (3a) 21476504025, (3b) 21545502813, (3c) 24127502235, (3d) 40702500019, (4a) 21512501331, (4b) 21492501653, (4c) 21492501685;
Recalling firm
CooperVision, Inc., West Henrietta, NY
Classification
Class II (temporary or reversible harm possible)
Status
Ongoing
Quantity
33990 lenses (17010 US, 16980 OUS)
Distribution
Domestic: CA, CO, FL, IL, KY, MI, NC, NE, NJ, NY, OH, OR, UT, VA, WA. International: Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Ireland, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Spain, Sweden, Switzerland, and United Kingdom.
Recall initiated
February 27, 2025
Classified
March 25, 2025
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1455-2025%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.

See every recall for this brand: Is CooperVision, Inc. recalled? →

Check this brand across all 8 databases →