Sekure Chemistry Sekisui Diagnostics Acetaminophen L3K Assay, In-vitro Diagnostic Catalog Number: 506-10: R1: 1 x 10ml, R2: 2 x 10ml Catalog Number: 506-30: R1: 3 x 10ml, R2: 6 x 10ml In-Vitro quantitative measurement of Acetaminophen in serum and plasma used m the diagnosis and treatment of acetaminophen overdose toxicity.
Class II · Terminated
Crystal formation in the Color Reagent (R2) when left on board automated analyzers causing samples to under recover
Advertisement
Advertisement
Affected lots / codes — check yours against thisCatalog Number: 506-10 Lot Codes 47677 47694 47649 47650 47662 47663 48253 48385 48386 Catalog Number: 506-30 Lot Codes: 47695 48294
- Recalling firm
- Sekisui Diagnostics P.E.I. Inc., Charlottetown,
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Terminated
- Quantity
- 14,407 kits.
- Distribution
- Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, IA, ID, IL, MA, , MN, MT and ND., and to the countries of Canada, France and Hong Kong.
- Recall initiated
- January 22, 2016
- Classified
- March 14, 2016
- Terminated
- August 9, 2016
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1171-2016%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1171-2016%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →