FDA Medical Device · Z-1171-2016

Sekure Chemistry Sekisui Diagnostics Acetaminophen L3K Assay, In-vitro Diagnostic Catalog Number: 506-10: R1: 1 x 10ml, R2: 2 x 10ml Catalog Number: 506-30: R1: 3 x 10ml, R2: 6 x 10ml In-Vitro quantitative measurement of Acetaminophen in serum and plasma used m the diagnosis and treatment of acetaminophen overdose toxicity.

Class II · Terminated

Crystal formation in the Color Reagent (R2) when left on board automated analyzers causing samples to under recover

Advertisement
Advertisement
Affected lots / codes — check yours against thisCatalog Number: 506-10 Lot Codes 47677 47694 47649 47650 47662 47663 48253 48385 48386 Catalog Number: 506-30 Lot Codes: 47695 48294
Recalling firm
Sekisui Diagnostics P.E.I. Inc., Charlottetown,
Classification
Class II (temporary or reversible harm possible)
Status
Terminated
Quantity
14,407 kits.
Distribution
Worldwide Distribution - US Distribution to the states of : AZ, CA, FL, IA, ID, IL, MA, , MN, MT and ND., and to the countries of Canada, France and Hong Kong.
Recall initiated
January 22, 2016
Classified
March 14, 2016
Terminated
August 9, 2016
Initiation
Voluntary: Firm initiated
Advertisement