FDA Medical Device · Z-1120-2026

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Class II · Ongoing

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Advertisement
Advertisement
Affected lots / codes — check yours against this400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0
Recalling firm
ICU Medical, Inc., Lake Forest, IL
Classification
Class II (temporary or reversible harm possible)
Status
Ongoing
Quantity
—
Distribution
Worldwide distribution - US Nationwide and the country of Philippines.
Recall initiated
December 15, 2025
Classified
January 16, 2026
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-1120-2026%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.