ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
Class II · Ongoing
Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
Advertisement
Advertisement
Affected lots / codes — check yours against thisSerial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
- Recalling firm
- Beckman Coulter, Inc., Indianapolis, IN
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Ongoing
- Quantity
- 92 units
- Distribution
- US Nationwide distribution.
- Recall initiated
- September 21, 2026
- Classified
- October 21, 2026
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0201-2027%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0201-2027%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots
above against your own product. This page reproduces public FDA data for readability;
EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →