FDA Medical Device · Z-0164-2027

TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.

Class II · Ongoing

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Advertisement
Advertisement
Affected lots / codes — check yours against thisCatalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Recalling firm
Howmedica Osteonics Corp., Mahwah, NJ
Classification
Class II (temporary or reversible harm possible)
Status
Ongoing
Quantity
605 trays
Distribution
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Recall initiated
September 16, 2026
Classified
October 15, 2026
Initiation
Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0164-2027%22
Advertisement
Important. Recalls target specific lots and codes — compare the affected lots above against your own product. This page reproduces public FDA data for readability; EveryRecall is independent and not affiliated with the FDA.
Check this brand across all 8 databases →