TS Universal Baseplate Finishing Tray, Triathlon. Catalog Number: 6543-6-506.
Class II · Ongoing
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Advertisement
Advertisement
Affected lots / codes — check yours against thisCatalog Number: 6543-6-506. GTIN: 07613327548082. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
- Recalling firm
- Howmedica Osteonics Corp., Mahwah, NJ
- Classification
- Class II (temporary or reversible harm possible)
- Status
- Ongoing
- Quantity
- 605 trays
- Distribution
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
- Recall initiated
- September 16, 2026
- Classified
- October 15, 2026
- Initiation
- Voluntary: Firm initiated