FDA Medical Device · Z-0028-2026

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Class I · Ongoing

Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

Advertisement
Advertisement
Affected lots / codes — check yours against thisModel Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.
Recalling firm
NOXBOX LTD, Sittingbourne,
Classification
Class I (most serious — risk of serious harm)
Status
Ongoing
Quantity
1667 units
Distribution
Domestic: TN
Recall initiated
September 9, 2025
Classified
October 6, 2025
Initiation
Voluntary: Firm initiated