NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Class I · Ongoing
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
Advertisement
Advertisement
Affected lots / codes — check yours against thisModel Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.
- Recalling firm
- NOXBOX LTD, Sittingbourne,
- Classification
- Class I (most serious — risk of serious harm)
- Status
- Ongoing
- Quantity
- 1667 units
- Distribution
- Domestic: TN
- Recall initiated
- September 9, 2025
- Classified
- October 6, 2025
- Initiation
- Voluntary: Firm initiated
Advertisement
Official source. This record is published by the U.S. FDA through the openFDA
enforcement database. View the unmodified government record:
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0028-2026%22
https://api.fda.gov/device/enforcement.json?search=recall_number%3A%22Z-0028-2026%22